Three-Dimensional Printed Flow Phantom Model of the Carotid Artery in Preterm Infants: Vessel Lumen Diameter Measurements Using Different Printing Materials

Authors

  • Sujith S. Pereira Neonatal Unit, Homerton University Hospital, Homerton Healthcare NHS Foundation Trust, London, UK and Centre for Genomics and Child Health, Blizard Institute, Barts and The London School of Medicine and Dentistry, London, UK https://orcid.org/0000-0001-7104-6872
  • Jonathan Reeves Clinical Physics, Royal London Hospital, Barts Health NHS Trust, London, UK
  • Malcolm Birch Clinical Physics, Royal London Hospital, Barts Health NHS Trust, London, UK and Centre for Neuroscience, Surgery and Trauma, Blizard Institute, Barts and The London School of Medicine and Dentistry, London, UK
  • Ahmed S Ali Neonatal Unit, Royal London Hospital, Barts Health NHS Trust, London, UK, Centre for Genomics and Child Health, Blizard Institute, Barts and The London School of Medicine and Dentistry, London, UK and Assiut University, Assiut, Egypt
  • Stephen T. Kempley Centre for Genomics and Child Health, Blizard Institute, Barts and The London School of Medicine and Dentistry, London, UK
  • Ajay K Sinha Neonatal Unit, Royal London Hospital, Barts Health NHS Trust, London, UK and Centre for Genomics and Child Health, Blizard Institute, Barts and The London School of Medicine and Dentistry, London, UK

DOI:

https://doi.org/10.14738/bjhr.1201.18129

Keywords:

3-dimensional, 3D printed flow phantom, carotid artery, preterm infant

Abstract

Background: Diameter forms an integral part of blood flow measurement. This study aimed to explore different three-dimensional (3D) printed materials to develop flow phantom models of the carotid artery in preterm newborn infants and to investigate best materials for diameter measurement validation. Methods: We produced a 3D printed Doppler flow phantom model with vessel lumen diameter of 2.0 mm with varying vessel characteristics using data from 21 preterm infants (right carotid vessel lumen diameter, wall thickness, blood flow measurements using Doppler ultrasound and distance of the carotid artery from skin surface) examined for research or clinical purposes. Flow phantom vessel lumen diameters were measured by a single operator blinded to flow phantom diameter. Results: 15 diameter measurements were performed. Ultrasound measured vessel lumen diameter measurements resulted in underestimation of the true lumen diameter. The measured mean (SD, range) diameter was 0.163 (0.105, 0-0.420) mm. This study found that difference in vessel lumen diameter measurements were least with the hybrid material (FLXA9895-DM) with shore value of 95 in matte finish. Vessel wall thickness was systematically overestimated in the majority of the measurements {Anterior wall thickness, mean(SD, range) 0.145(0.081, 0.020–0.300) mm and posterior wall thickness, mean(SD, range) 0.103(0.117,  minus 0.100–0.370) mm}. Conclusion: We successfully produced a 3D printed flow phantom model of the carotid artery in preterm infants with varying vessel characteristics and identified flow phantoms that produced the least difference in ultrasound measured vessel lumen diameter measurements. Researchers and clinicians can use this information for further studies involving ultrasound diameter measurements.

References

Filippou V, Tsoumpas C. Recent advances on the development of phantoms using 3D printing for imaging with CT, MRI, PET, SPECT, and ultrasound. Med Phys. 2018.

Gross BC, Erkal JL, Lockwood SY, Chen C, Spence DM. Evaluation of 3D printing and its potential impact on biotechnology and the chemical sciences. Anal Chem. 2014;86(7):3240-53.

Domingo-Roca R, Asciak L, Windmill JFC, Mulvana H, Jackson-Camargo JC. Non-destructive Analysis of the Mechanical Properties of 3D-Printed Materials. J Nondestr Eval. 2022;41(1):22.

Nikitichev DI, Barburas A, McPherson K, Mari JM, West SJ, Desjardins AE. Construction of 3-Dimensional Printed Ultrasound Phantoms With Wall-less Vessels. Journal of ultrasound in medicine : official journal of the American Institute of Ultrasound in Medicine. 2016;35(6):1333-9.

Pereira S, Reeves J, Birch M, Ali A, Sinha A, Kempley S. Diameter measurements in three-dimensional printed flow phantom model of the carotid artery in preterm infants. Journal of Biomedical Engineering and Medical Imaging. 2021;8(6):8-21.

Pereira S, Reeves J, Birch M, Ali A, Sinha A, Kempley S. Three-dimensional printed flow phantom model of the carotid artery in preterm infants for training and research. Journal of Biomedical Engineering and Medical Imaging. 2021;8(6):102-14.

Pereira SS, Reeves J, Birch M, Ali AS, Sinha AK, Kempley ST. Inter- and intra-scanner differences in Doppler flow measurements using three-dimensional printed flow phantom model of the carotid artery in preterm infants. Journal of Biomedical Engineering and Medical Imaging. 2022;9(1).

Pereira S, Reeves J, Birch M, Finton-James S, Verma K, Krug R, et al. A realistic flow phantom model of the carotid artery in preterm infants for training and research. Ultrasound. 2020;28(3):145-54.

Pereira SS, Sinha AK, Morris JK, Wertheim DF, Shah DK, Kempley ST. Blood pressure intervention levels in preterm infants: pilot randomised trial. Arch Dis Child Fetal Neonatal Ed. 2018.

Teirlinck CJ, Bezemer RA, Kollmann C, Lubbers J, Hoskins PR, Ramnarine KV, et al. Development of an example flow test object and comparison of five of these test objects, constructed in various laboratories. Ultrasonics. 1998;36(1-5):653-60.

Harmsen SS. Acoustic impedance matching using 3D printed lenses. Netherlands: University of Twente; 2018.

Downloads

Published

2025-02-25

How to Cite

Pereira, S. S., Reeves, J., Birch, M., Ali, A. S., Kempley, S. T., & Sinha, A. K. (2025). Three-Dimensional Printed Flow Phantom Model of the Carotid Artery in Preterm Infants: Vessel Lumen Diameter Measurements Using Different Printing Materials. British Journal of Healthcare and Medical Research, 12(01), 153–163. https://doi.org/10.14738/bjhr.1201.18129